MC100

Analogue Multicare Controller

MC100

The MC100 has been specifically designed for use with the Apollo Lighting Multicare luminaire equipped with analogue dimming control gear in order to meet the three lighting level requirements necessary for overbed illumination.

  1. 'Patient Reading' appropriate for day to day use
  2. 'Inspection' lighting
  3. 'Examination' lighting

Suitable for flush or surface mounting installation as required.

Operation

  • To achieve patient reading lighting as described above, turn on the left switch
  • To achieve inspection lighting, turn on the right switch
  • To achieve examination lighting, turn on both the left switch and the right switch and turn left rotary controller fully clockwise

Order Code

  • MC100

Dimensions

MC100 Dimensions
W H
90 90

Technical

Rotary Switch Contacts1A resistive rating, Note: Only to be used to switch one ballast
Rotary Switch Terminals4 screw terminals (two for the 1–10V DC and two for the 240V mains switch)
the 1–10V DC control cable should be mains rated and run separately or screened from the mains cables
White Mains Switch Contacts20A nominal rating, single gang, single pole, two way
White Mains Switch Terminals4.0mm² conductors
MountingSuitable for installation on 72 x 72 x 41mm deep single gang back box (not supplied)

Wiring

MC100 Wiring Diagram